Next: Conclusion
Up: Limits of ultrashort pulse
Previous: Nature of the stability
Pulse stabilization
In this section we summarize the ways to achieve the stable single
pulse operation in the system prone to the multiple pulse lasing.
As the shortest pulse width for the fixed GDD is achieved by the
growth, the main goal of the system's optimization is to
enhance the stability against the bounded perturbations. The
obvious way in this direction is the decrease of the intracavity
energy. Such decrease can be undesirable for some systems (the
high-power pulse sources, for example) and can degrade the
self-start ability for the Kerr-lens mode-locked lasers. Moreover,
this decrease can reduce the gain saturation and enhance the
destabilization due to the background growth.
Let us describe the main approaches to the single pulse
stabilization.
1) The increase of the modulation depth, i. e. the
parameter, expands the region of the single pulse generation with
subsequent pulse shortening and reduces the zone of unstable
operation (transition from b to a in Fig.
). It should be noted, that such scenario decreases the
self-starting ability, and may be a bad choice for many
solid-state lasers. However, this way is acceptable for the Cr
:ZnSe laser due to its large values of
and
(for given pump intensity
).
2) The increase of the linear loss, i. e.
, expands the
stability region for the single pulse operation. However, the
pulse shortening that can be achieved in this way is comparatively
small (compare Fig.
, a and Fig.
,
c). This is so because the background amplification is
not suppressed in this case owing to the reduced gain saturation
(pulse energy is lower in the case of higher
), so that
there is only small improvement to the minimum GDD defining the
minimal pulse duration.
3) The most appropriate choice is the increase of the
parameter (Fig.
, d in comparison to Fig.
, a). The increase of the gain saturation in
comparison to the self-phase modulation contribution prevents both
the background and bounded perturbation growth. Unfortunately,
this parameter is constant for the given laser medium with fixed
length x. Taking shorter active medium with the same
enhances the stability. In this context, the Cr
:ZnSe is very attractive in comparison to other media due
to the large value of
. For example:
=1.3
,
=6.9
,
=6.3
,
=3.6
The high value of
for ZnSe is partially compensated by
the large
,
and n (see
definition of
in section
).
4) Another approach is to decrease
by decreasing
or
(Fig.
, e in comparison to
Fig.
, a) or to decrease
(Fig.
, f). Unfortunately, the advantage of Cr
:ZnSe consisting in its large absorption cross-section
turns into disadvantage: other parameters being equal, the larger
and smaller
result in the larger
. For example, for 1.5 W of the absorbed pump power at the
center of the corresponding absorption lines and 100
m
diameter of the pumping beam
=
5.9
,
= 6.3
,
= 5.4
,
=
1.2
. The upper-laser level lifetime
is the
material constant, and can be reduced (e.g. by heating or
concentration quenching) only at the expense of higher laser
threshold.
It should be noted, that according to Fig.
, the pulse
intensity decrease provided by the methods 2) - 4) increases value
of
, at which the minimum pulse duration is achieved.
This can demand a very thorough optimization of the laser design.
Note also that the finger-like shape of the stability region for
the small negative GDD providing the minimal pulse durations
requires fine system optimization by the appropriate choice of
. This feature is more critical for higher
(compare graphs a and b in Fig.
).
Next: Conclusion
Up: Limits of ultrashort pulse
Previous: Nature of the stability
V.L. Kalashnikov
2002-12-28