FNV-200/CNV-100 Falling / Controlled Needle Viscometer
Download
FNV-200/CNV-100 Datasheet In Adobe PDF Format
The
Falling Needle Viscometer (FNV) consists of a metal or glass needle with
hemispherical ends falling vertically in a fluid with its longitudinal axis
along the axis of a cylindrical container. Also, the Controlled Needle
Viscometer (CNV) consists of a metal needle with hemispherical ends and its top
end is connected to the extension bar plus the weight holder. The CNV is the
same as the FNV except for the needle dropped. A measurement of the needle
terminal velocity allows a determination of the viscous properties and density
of the test fluid. The falling time is measured automatically by means of Hall
sensors and a magnet in the needle. The CNV comes with a small sample insert kit
for those cases when the sample supply is limited.
Key Advantages:
- Measures Absolute Viscosity
without Instrument Calibration
- Better than One Percent Accuracy
and Repeatability
- Large Viscosity Range: 0.1 to 108 cP
- Accurate Temperature Control
- Large Temperature Range of Fluids
- Simultaneous Density and Viscosity
Measurements
- Thermal Expansion Coefficient
- Relative Sedimentation Rates
- Automatic Falling Time
Measurements to 0.001 s
- Microprocessor-based Console with
RS-232 Interface
Specifications for FNV-200
Falling Needle Viscometer
| Viscosity Range |
0.5 to 2.4x106 cP
(mPa s)* |
| Shear Rate Range |
Approx. 10-4 to
104 1/s** |
| Temperature Range |
-40 to 150oC* |
| Accuracy & Repeatability |
Better than 1% |
| Test Sample Volume |
Approx. 110cm3* |
| Needle Material |
Metal (Al, Ti, S.S.) or
Glass |
| Needle Densities |
1.01 to 8 g/cm3* |
| Total Instrument Weight |
18.2 kg (40 lbs) |
| Size |
O.A.H. 55cm, O.A.W. 36cm |
*outside this range
consult SBS.
**depends on the sample and system.
Specifications for CNV-100
Controlled Needle Viscometer
Metal (Ti or Al) Needle Kit
| Viscosity Range |
0.1 to 30 cP (mPa s)* |
| Accuracy & Repeatability |
Better than 1% |
| Test Sample Volume |
Approx. 85cm3* |
Small Sample Insert Kit
| Viscosity Range |
3 to 108 cP (mPa
s)* |
| Accuracy & Repeatability |
Better than 1% |
| Test Sample Volume |
Approx. 7cm3* |
*outside this range
consult SBS. |