1) Lower Buffer core into the lower buffer chamber so that the negative electrode fits into the opening in the gold plate.
2) Insert gel tension wedge behind the buffer core. Make sure the wedge is in its unlocked position. It should rest on the bottom of the lower buffer chamber.
3) Remove gel cassette from pouch by cutting with scissors, drain packing buffer, rinse with deionized water.
4) Peel off tape covering slot on back of gel cassette.
5) In one fluid motion, pull comb out of cassette.
6) Insert gel cassettes into the lower buffer chamber. Place one cassette behind the core and one cassette in front of the core. For each cassette, the shorter "well" side of the cassette faces in towards the buffer core. The slot on the back of the cassette must face out towards the lower buffer chamber.
7) If you are running only one gel, replace the rear gel cassette with the buffer dam.
8) Pull forward on the gel tension lever in a direction towards the front of the Xcell SureLock unit until lever comes to a firm stop and the gels appear snug against the buffer core.
9) Fill the upper buffer chamber (between the two gels) to the top of the gel cassettes with 1X Running Buffer. Ensure that the chamber is not leaking by observing level. If the level of running buffer drops, the electrophoresis core and cassettes are not properly seated.
Fill the lower buffer chamber with 200 ml of the 1X running buffer.
10) Use a loading pipette to gently wash the cassette wells with the running buffer in the upper chamber.