Diagnostic Imaging 5.19.99
Differential Absorption
- Maximal differential absorption results from lower kVp values
- But as the kVp is lowered patient dose increases
- AS you drop the kVp the film becomes much lighter due to weaker beam, this is compensated by increasing the mA and the time.
- At lower energies the % of photoelectric interactions increases (relative to compton), above 40 kVp compton is the predominant form of radiation.
- To image small differences in soft tissue use low kVp technique ( for maximum differential absorption)
- Four things that can be seen on an X-ray is air, fat, water, metal
- The interaction of X-rays and tissue is proportional to the density (specific gravity) of the tissue.
- Attenuation is the reduction in the number of X-rays as they penetrate matter. X-rays are attenuated exponentially.
X-Ray Quantity- The numbber of x-ray photons in a beam
- Quantity=output intensity=radiation exposure
- Roentgens measure of photons often stated in milliroentgns (mR)
- International community uses gray (Gy) and Sievert (Sv)
- Gray is patient dose
- Sievert is occupational exposure both are the same numerically
- In diagnostic radiology 1 rad = 1 rem = 1 roentgen
- The final measure is in terms of mR/mAs
- The high frequency generators produce double the mR/mAs over the single phase machine when the kV is the same on both.
- At 70 kVp there is usually 3-6 mR/mAs
- As the time goes up the film gets darker, that’s why the high frequency generators require less time.
- MAs
- Quantity is directly proportional
- If the mAs doubles the film density double or gets darker
- If the mAs is halved the film density is halved, gets lighter.
- KVp
- 15% rule not in book
- an increase in the kVp by 15% must be balanced by a 50% reduction in mAs
- 80kVp X 15% = 92
- Conversely a decrease in kVp by 15% must be balanced by a doubling of the mAs
- The result in either case are two films of equivalent average density but a change in contrast.
- Distance
- Inverse square law
- To maintain the density of the film if the FFD is increased 100% (e.g 40"-80") you need to quadruple the mAs.
- Conversely if you cut the FFD by 50% (e.g 80"-40") you would need one 1/4th the mAs to create films of equal density.
- If you increase the distance by 50% (40"-60") you would double the mAs to create films of equal density.
- If the FFD is increased from 40"-72" the mAs is tripled
- Inverse Square Law = I1/I2= (d2/d1)^2
- Example: A technique at 40" is exposed at 300mr. What would the exposure be if the tube was moved to 72" w/out changing the kVp or mAs.à
I1= 300mr(40"/72")^2 I1= 93mr
- Example 2: A technique at 72" is exposed to 300mr . What would the exposure dose be if the tube was moved to 40".à
I1= 300mr(72"/40")^2 I1=972mr
- With tube tilt you move the tube closer by 1 inch for every 5 degrees of tilt. I1= 15mr(43"/40")^2 I1= 13mr which is a 14% decrease, therefore a tube tilt of 15 degrees does not need to be moved however, above 15 degrees the tube must be moved closer.
X-Ray Quality
- Refers to the penetrating power of the beam and evaluating the kVp
- Hard vs. Soft X-rays
- Hard x-rays have higher kV values
- Soft x-rays have lower kV values
- Measured by half valued layer (HVL)
- HVL is the thickness of an absorbing material necessary to reduce the intensity to ½ of its original value.
- As kVp goes up so does the HVL thickness
- Factors affecting beam quality
- KVp
- Filtration
- Types of filtration
- Inherent- oil and glass and plastic plug built into tube
- Added- Government stipulated thickness of aluminum needed to a filter
- Compensating- Slabs of aluminum and reduces the beam at the top of the thorasic spine.
Beam Restricting Devices
- Two types of photons interact with the film to form image: those which don’t interact and Compton scatter x-rays
- Remenant x-rays- photons that exit the patient and interact with the film
- Collimation reduces patient dose by:
- Limiting the volume of the tissue irradiated
- Secondarily reducing the amount of scatter (also reduces fogging)
- Failure to collimate is the greatest contributor to needless patient exposure.
- Factors affecting scatter radiation
- As kVp increases the relative number of compton interactions increases.
- In general high (optimal) kVp technique is preferred to low kVp technique due to patient dose considerations.
- Field size- as the area of the beam enlarges, scatter increases. Seen more on small film sizes.
- For a spot film, the mAs must be increased to keep the density the same due to the scatter.
- Patient thickness- the thicker the body part the more scatter.
- Patient thickness is under your control
- Deciding to perform lumbar spine and abdominal films in the recumbent versus upright positions
- If you decide to use patient compression
Control of scatter radiation
- Use beam restriction and grids
- Beam restriction
- Aperture diaphragm
- Cones or cylinders
- Variable aperture collimator
- Aperture diaphragms are generally not suitable for general practice
- Cones are primarily used by dentists
- General practice requires variable aperture collimator.
- Variable aperture collimators (positive beam limitation)
- Off-focus radiation may be reduced by a first stage entrance shuttering device
- Second stage collimator shutter leaves are made of lead (at least 3mm thick), work in pairs, independently, yielding square or rectangular fields, which compliment the film size being used.
- A light localization by lamp and mirror.
- Place rheostat on x-ray room lights
- Collimator bulbs are expensive, handle with tissue when installing or their life will be shortened and crack.
- Light must coincide with the x-ray beam.
- If the equipment is properly installed and maintained the resulting film will show evidence of four sides of collimation.
- Collimator adds about 1mm of aluminum filtration equivalent